Detección de gases y vapores inflamables y tóxicos
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

// DetecciÓn de gases y vapores inflamables y tÓxicos - Medidores Portables - Un gas - FORMAL DEMETER

FORMALDEMETER htV
Portable Direct Reading Formaldehvde Monitor

Formaldemeter htV
Portable Direct Reading Formaldehyde Monitor

• Easy to use – minimal training required
• Rapid response time
• Real time readings allow immediate response to results
• Reliable electrochemical sensor
• Does not require inconvenient ancillary sampling equipment or readers
• Simple in-field calibration
• Small, hand-held instrument
• Light-weight
• Optional Base Unit for continuous monitoring

 

The Formaldemeter htV offers a simple way to monitor levels of airborne formaldehyde in the environment. The instrument is easy to use with a simple four-button keypad and features an easy-to-read four-digit display. It employs a well-proven electrochemical formaldehyde sensor that will provide an immediate indication of formaldehyde levels.
Unlike other formaldehyde monitoring devices, such as, color stain tubes and badges, the Formaldemeter htV is capable of measuring many samples consecutively without the need for inconvenient ancillary equipment. Real-time results also mean that potential overexposures can be identified as they occur, without the delay and expense associated with laboratory analysis.

Also available, the Glutaraldmeter 3, for glutaraldehyde monitoring.


Specifications Applications


Sensor: Electrochemical
Calibration: In-field formaldehyde calibration standard
Display: 4-digit LCD
Sampling Frequency: Typically once every 3 to 5 minutes depending on previous reading
Sample Volume: Approx. 10 ml
Detection Range: 0.05-10 ppm (optional high range: 80 ppm)
Precision: 10% at 2 ppm level
Response Time: <10 seconds
Temperature: By key interrogation, displayed in degrees C
Humidity: By key interrogation, displayed in % rH
Power: 9 volt alkaline battery
Enclosure: ABS plastic
Weight: Instrument: 10 ounces (310g)
Complete kit: 1.75 lbs. (795g)
Dimensions: Instrument: 6 x 3.25 x 1.25 inches (150 x 80 x 32 mm)
Carrying Case: 10.5 x 8.5 x 2 inches (265 x 215 x 50 mm)

Cross Sensitivity Chart


Formaldehyde is a common contaminant found in new construction, residential housing and commercial buildings following renovation. It is used in many products, industries and other types of activities, including: Textile treatment, Paper products, Medical labs, Biological, Laminated boards, Dye stuffs
Pharmaceuticals, Particle boards, Synthetic resins, Commercial buildings, Horticulture, Cosmetics, Paints, Residential construction, Mortuaries, Hospitals

Compound

Acetone
Acetaldehyde
Acetic acid
Ammonia
Butanol
Carbon monoxide
Ethylene
Ethanol
Formaldehyde
Glutaraldehyde
Methanol
Phenol
Propanol
Resorcinol

ppm = 1 ppm reading

No Response
8-12
No Response
71000
Note (2)
100
160
12-20
1.0
7100 at 25° C, Note (3)
50
5, Note (1)
Note (2)
5, Note (1)

Notes:
(1) Completely removed with filter provided
(2) Interference above 60° C nd 20 ppm
3) Monitor with Glutaraldmeter 3


Accessories and Replacment Parts

Phenol Filters (supplied with instrument kit)
These filters can be used, when necessary, for the removal of phenol and resorcinol.

Calibration Standard (supplied with instrument kit)
The calibration standard is a tube wich produces a known concentration of formaldehyde vapo for testing and, if necessary, adjusting the accuracy of the monitor. The standard has a 100-sample capacity or a useful life of six months, whichever occurs sooner.

Options

AMS-2 Base Unit
The Base Unit automates operation of the Formaldemeter htV for semi-continuous sampling.
Monitoring of all three parameters (formaldehyde, temperature and humidity) is possible with the AMS-2, giving real-time data and alarm (TWA/STEL) levels. A built-in printer provides a record of the results. All data is stored and can be downloaded to a PC via RS-232 and available download software.

Options

AMS-2/htV Download Software
Data logged into the AMS-2 can be viewed on the LCD and outputted to the integral printer.
If required, the data can be downloaded to most PC based spreadsheets using the AMS-2/htV
Download Software.

Ordering Information

02700-100
Instrument kit, with Formaldemeter htV, 10 phenol filters, thermometer, pen, calibration standard, instruction manual and case.
02710-000
Instrument kit, with Glutaraldmeter 3, 10 phenol filters, thermometer, pen, calibration standard, instruction manual and case.
02700-251
Optional data logging station, Model AMS-2 Base Unit for htV
02700-256
Optional PC download software for AMS-2/htV
02700-001
Formaldehyde Calibration Standard, typical life 6 to 12 months
02710-001
Glutaraldehyde Calibration Standard, typical life 6 to 12 months
02700-002
Replacement Phenol Filters, pack of 10
02700-003
Replacement Formaldehyde Sensor, typical life 3-4 years


. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
 